Journal

The Journal of the Japanese Society for Clinical Microbiology

Biblioraphy Information

[Vol.25 No.2 contents]
Japanese / English

download PDF Full Text of PDF (208K)
Article in Japanese

ArticleTitle Evaluation of matrix-assisted laser desorption ionization-time of flight mass spectrometry systems for identification of streptococci
Language J
AuthorList Yusuke Ota1), Takehisa Matsumoto2), Eriko Kasuga2), Kazuki Horiuchi2), Tatsuya Negishi2), Tomomi Yaguchi2), Tatsuya Natori2)
Affiliation 1) Department of Health and Medical Sciences, Shinshu University, Nagano, Japan
2) Department of Laboratory Medicine, Shinshu University Hospital, Nagano, Japan
Publication J.J.C.M.: 25 (2), 111-116, 2015
Received October 1, 2014
Accepted December 2, 2014
Abstract The performance of matrix-assisted laser desorption ionization-time of flight mass spectrometry systems were evaluated in the identification of streptococci. A total of 135 clinical isolates were analyzed using a Bruker MALDI Biotyper system. This system correctly identified 100% (135/135) of Streptococcus isolates to the genus-level and 81.5% (110/135) to the species-level for two sample preparation methods (on-plate formic acid preparation and ethanol-formic acid extraction). Several bacterial species could not be identified to the species-level by this system because of few number of reference spectra in current reference library. It is necessary to improve reference library in MALDI Biotyper for the more accurate identification of clinically significant Streptococcus strains.
Keywords
Copyright © 2002 The Japanese Society for Clinical Microbiology